kestrel4000.com.au

  • Increase font size
  • Default font size
  • Decrease font size
Home Specification
Kestrel 4000 Specification
Measurement
Response Time
UnitsOperational RangeResolutionAccuracy (+/-)Specification Range
Wind Speed
1 second
m/s0.4 to 60.0 m/s0.1Larger of 3% of reading or least significant digit0.4 to 40.0 m/s
ft/min59 to 11,948 ft/min159 to 7877 ft/min
km/h1.0 to 218.0 km/h0.11.0 to 144.0 km/h
mph0.8 to 135.0 mph10.8 to 89.0 mph
knots0.6 to 118.3 kt0.10.6 to 78.0 kt
Beaufort0 to 12 B0.10 to 12 B
1 inch diameter impeller with precision axle and sapphire bearings. Off-axis accuracy -1% @ 5º off-axis; -2% @ 10º; -3% @ 15º. Calibration drift < 1% after 100 hours use at 16 MPH / 7 m/s. Sustained operation above 60 MPH / 27 m/s will wear impeller rapidly and may cause destruction of impeller. Replacement impeller (NK PN-0801) may be field-installed without tools (US Patent 5,783,753).
Temperature
1 second
°F-49.0 to 257.0 °F0.11.8 °F-20.0 to 158.0 °F
°C-45.0 to 125.0 °C0.11.0 °C-29.0 to 70.0 °C
Measures air, water and snow temperature. Thermally isolated, hermetically sealed, precision thermistor mounted externally (US Patent 5,939,645). Calibration drift negligible.
Relative Humidity
1 minute
%RH0.0 to 100.0 %0.13.0 %RH5.0 to 95.0 % non-condensing
Polymer capacitive humidity sensor mounted in thin-walled chamber external to case for rapid, accurate response (US Patent 6,257,074). (To achieve stated relative humidity accuracy, unit must be permitted to equilibrate to external temperature when exposed to large, rapid temperature changes and must be kept out of direct sunlight.) Calibration drift +/- 2% over 24 months. Relative humidity may be recalibrated at factory or in field using Kestrel Humidity Calibration Kit (NK PN-0802).
Pressure
1 second
(mb & PSI 4000 model only)
inHg0.3 to 32.5 inHg0.010.05 inHgAt 77.0 °F, 22.1 to 32.5 inHg
hPa / mb10.0 to 1100.0 hPa / mb0.11.5 hPa / mbAt 25.0 °C, 750 to 1100 hPa / mb
PSI0.15 to 16.0 PSI0.010.02 PSIAt 77.0 °F, 10.9 to 16.0 PSI
Monolithic silicon piezoresistive pressure sensor with second-order temperature correction. Maximum error over temperature range 32 to 158 °F (0 to 70°C) , +/- 0.06 inHg / +/-2.0 hPa. Calibration drift typically -0.03 inHg / -1.0 hPa per year. Pressure sensor may be recalibrated at factory or in field.
Altitude
1 second
ft-6000 to 30000 ft150 ftAt 77.0 °F, <19,700 ft. Max error +/- 98 ft
m-2000 to 9000 m115 mAt 25.0 °C, <6,000 m. Max error +/- 30 m
Temperature compensated pressure (barometric) altimeter.
Wind Chill
1 second
°F0.7 to 135.0 MPH, -49.0 to 257.0 °F0.11.8 °F1.8 to 89.0 mph, -50.0 to 50.0 °F
°C0.4 to 60.0 m/s, -45.0 to 125.0 °C0.11.0 °C0.4 to 40 m/s, -45.6 to 10.0 °C
Calculated from the primary measurements of wind speed and temperature. Utilizes the NWS Wind Chill Temperature (WCT) Index, revised 2001, with wind speed adjusted by a factor of 1.5 to yield equivalent results to wind speed measured at 10 m above ground. (Specification temperature limits established by WCT Tables.)
Heat Index
1 minute
°F0.0 to 100.0 %RH, -49.0 to 257.0 °F0.13.6 °F70.0 to 130.0 °F, 0 to 100% RH
°C0.0 to 100.0 %RH, -45.0 to 125.0 °C0.12.0 °C21.1 to 54.4 °C, 0 to 100 %RH
Calculated from the primary measurements of temperature and relative humidity. Utilizes the NWS Heat Index (HI) tables. (Specification temperature limits established by HI tables.)
Dewpoint
1 minute
°F0.0 to 100.0 %RH, -49.0 to 257.0 °F0.13.6 °F-20.0 to 158.0 °F, 20.0 to 95.0% RH
°C0.0 to 100.0 %RH, -45.0 to 125.0 °C0.12.0 °C-29.0 to 70.0 °C, 20.0 to 95.0 %RH
Calculated from the primary measurements of temperature and relative humidity. Temperature to which the air would need to be cooled at a constant pressure to become saturated.
Wet Bulb Temperature 1 minute °F-49.0 to 257.0 °F, 0.0 to 100.0 %RH, 8.86 to 32.48 inHg0.13.6 °F32.0 to 100.0 °F, 5.0 to 95.0% RH, 8.86 to 32.48 inHg, <19700 ft
°C-45.0 to 125.0 °C, 0.0 to 100.0 %RH, 300.0 to 1100.0 hPa0.12.0 °C0.0 to 37.8 °C, 5.0 to 95.0 %RH, -2000.0 to 9000.0 hPa, <6000 m
Calculated from the primary measurements of temperature, relative humidity and pressure. Temperature indicated by a wet bulb psychrometer.
Density Altitude
1 second
ft-49.0 to 257.0 °F, 0.0 to 100.0 % RH, 8.86 to 32.48 inHg124632.0 to 100.0 °F, 5.0 to 95.0 %RH, 8.86 to 32.48 inHg, <19700 ft
m-45.0 to 125.0 °C, 0.0 to 100.0 %RH, 300.0 to 1100.0 hPa1750.0 - 37.8 °C, 5.0 to 95.0 %RH, -2000 to 9000 hPa, <6000 m
Calculated from the primary measurements of temperature, relative humidity and pressure. Air density converted to equivalent sea level elevation at the International Standard Atmosphere.
Max/Avg Wind Speed (Air Velocity),
Crosswind, Headwind/Tailwind
One-button clear and restart of Max Wind Gust and Average Wind measurement.
Data Storage / DisplayMinimum, maximum, average and logged history stored and displayed for every measured value. 2000-point data logger with graphical display. Auto data storage; interval settable from 2 seconds to 12 hours. Manual data capture.
Data UploadRequires optional PC interface (NK PN-0830) and provided software. RS-232 connection with USB adapter available.
DisplayMultifunction, multi-digit programmable dot-matrix display.
Display Update1 second.
Display BacklightChoice of aviation green or visible red electroluminescent backlight. Automatic or manual activation.
Clock / CalendarReal-time hours:minutes:seconds clock, calendar, automatic leap-year adjustment.
Operational Temperature Range
(LCD and Batteries)
The operational temperature range of the liquid crystal display and batteries is 14° F to 131° F / -10 °C to 55 °C. Beyond the limits of the operational temperature range, the unit must be maintained within range and exposed for minimum time necessary to take reading.
Storage Temperature-22 °F to 140 °F / -30 °C to 60 °C.
Auto ShutdownUser-selectable: 15 or 60 minutes with no key presses or disabled.
LanguagesEnglish, French, German, Italian, Spanish.
CertificationsCE certified. Individually tested to NIST-traceable standards (written certificate of tests available at additional charge).
BatteriesAAA Alkaline, two, included. Average life, 400 hours of use, +/-depending on backlight use.
EnvironmentalWaterproof (IP67 standard). Drop-tested (MIL.STD.810F; unit only. Substantial impact may damage replaceable impeller.).
DimensionsUnit 5.0 x 1.8 x 1.1 in / 12.7 x 4.5 x 2.8 cm.
WeightUnit 3.6 oz / 102 g.